It is intended for properly prepared instruments in dental, medical and other approved professional settings.
Class B cycles are designed for wrapped and unwrapped solid, porous and hollow instruments, in accordance with the manufacturer's instructions. The autoclave is not intended for sterilizing liquids.
Features and benefits
- LCD screen displaying time, temperature, pressure and cycle progress.
- Built-in thermal printer for documenting sterilization cycles.
- USB connection for saving cycle records.
- Ten automatic programs: seven sterilization cycles, two test cycles and one cleaning program.
- User settings, preheating and temperature-hold options.
- Clean and waste-water tanks, self-diagnostics and dual door locking.
Example sterilization programs
- 121 °C, solid instruments · Pressure: 110 kPa · Cycle time after heating: 23 min
- 121 °C, porous instruments · Pressure: 110 kPa · Cycle time after heating: 27 min
- 121 °C, hollow instruments · Pressure: 110 kPa · Cycle time after heating: 30 min
- 134 °C, solid instruments · Pressure: 210 kPa · Cycle time after heating: 7 min
- 134 °C, porous instruments · Pressure: 210 kPa · Cycle time after heating: 11 min
- 134 °C, hollow instruments · Pressure: 210 kPa · Cycle time after heating: 14 min
- 135 °C, prion cycle · Pressure: 210 kPa · Cycle time after heating: 29 min
Specifications are based on supplier information. Before use, consult the manufacturer's instructions and required sterilization procedures.